DIAGnostic:SERVice:NSOurce <State>
This command turns the 28 V supply of the BNC connector labeled [noise source
control] on the R&S FSV/A on and off.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
DIAG:SERV:NSO ON
Manual operation: See "Noise Source Control" on page 352
OUTPut<up>:IF:IFFRequency <Frequency>
This command defines the frequency for the IF output of the R&S FSV/A. The IF fre-
quency of the signal is converted accordingly.
This command is available in the time domain and if the IF output is configured for IF.
Suffix:
<up>
.
Parameters:
<Frequency> *RST: 50.0 MHz
Default unit: HZ
Manual operation: See " IF Out Frequency " on page 353
OUTPut<up>:IF:STATe <State>
Enables or disables output of the measured IF value at the "IF" output connector.
Suffix:
<up>
.
1..n
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0