Loading...
Network and Remote Operation
R&S
®
FSVA3000/ R&S
®
FSV3000
707User Manual 1178.8520.02 ─ 01
"VXI-11 Dis-
covery"
If enabled, connected devices are detected automatically using the
VXI-11 protocol (see "VXI-11 Protocol" on page 647)
Remote command:
SYSTem:LXI:INFO? on page 1125
LXI Password
Password for LAN configuration. The default password is LxiWebIfc.
Remote command:
SYSTem:LXI:PASSword on page 1126
LXI Manufacturer Description
Instrument description of the R&S FSV/A
Remote command:
SYSTem:LXI:MDEScription on page 1126
LAN Reset
Resets the LAN configuration to its default settings (LCI function).
According to the LXI standard, an LCI must set the following parameters to a default
state.
Parameter Value
TCP/IP Mode DHCP + Auto IP Address
Dynamic DNS Enabled
ICMP Ping Enabled
Password for LAN configuration LxiWebIfc
Remote command:
SYSTem:LXI:LANReset on page 1126
13.5.5 Remote Errors
Access: [SETUP] > "Network + Remote" > "Remote Errors " tab
The error messages generated by the R&S FSV/A during remote operation are dis-
played here.
The messages are displayed in the order of their occurrence; the most recent mes-
sages are placed at the top of the list.
Network and Remote Control Settings