EasyManua.ls Logo

R&S FSV3000 Series - Page 1125

R&S FSV3000 Series
1268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1125User Manual 1178.8520.02 ─ 01
SYSTem:ERRor:DISPlay <State>
This command the error display during remote operation on and off.
If activated, the R&S FSV/A displays a message box at the bottom of the screen that
contains the most recent type of error and the command that caused the error.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
SYST:ERR:DISP ON
Manual operation: See "Display Remote Errors" on page 701
SYSTem:IDENtify:FACTory
This command resets the query to *IDN? to its default value.
Manual operation:
See "Reset to Factory String" on page 700
SYSTem:IDENtify[:STRing] <String>
This command defines the response to *IDN?.
Parameters:
<String> String containing the description of the instrument.
Manual operation: See "Identification String" on page 700
SYSTem:KLOCk <State>
This command activates the local lockout (remote control) or returns to the local mode.
Parameters:
<State> ON
LLO (local lockout)
OFF
GTL (go to local)
*RST: OFF
Example:
SYST:KLOK ON
Activates LLO (remote control)
Manual operation: See "Local" on page 709
SYSTem:LXI:INFO?
This command queries the LXI settings.
Return values:
<LXIInfo> <current version> | <LXI class> | <Computername> |
<MAC adress> | <IP adress> | <Auto MDIX>
String containing the current LXI parameters.
Configuring the R&S FSV/A