Common Analysis and Display Functions
R&S
®
FSVA3000/ R&S
®
FSV3000
486User Manual 1178.8520.02 ─ 01
9.5 Trace Configuration
A trace is a collection of measured data points. The trace settings determine how the
measured data is analyzed and displayed on the screen.
Standard Traces....................................................................................................486
Spectrograms........................................................................................................498
Trace Math............................................................................................................ 516
9.5.1 Standard Traces
9.5.1.1 Basics on Setting up Traces
Some background knowledge on traces is provided here for a better understanding of
the required configuration settings.
Mapping Samples to sweep Points with the Trace Detector.................................486
Analyzing Several Traces - Trace Mode............................................................... 489
How Many Traces are Averaged - Sweep Count + Sweep Mode.........................490
How Trace Data is Averaged - the Averaging Mode............................................. 491
Trace Smoothing...................................................................................................491
Mapping Samples to sweep Points with the Trace Detector
A trace displays the values measured at the sweep points. The number of samples
taken during a sweep can be much larger than the number of sweep points that are
displayed in the measurement trace.
Example:
Assume the following measurement parameters:
Sample rate: 32 MSamples / s
sweep points: 1000
sweep time: 100 ms
Span: 5 GHz
During a single sweep, 3.2 * 10
6
samples are collected and distributed to 1000 sweep
points, i.e. 3200 samples are collected per sweep point. For each sweep point, the
measured data for a frequency span of 5 MHz (span/<sweep points>) is analyzed.
Note that if you increase the number of sweep points, the frequency span analyzed for
each point in the trace decreases, making the result more stable.
See also Chapter 8.5.1.8, "How Much Data is Measured: Sweep Points and Sweep
Count", on page 381.
Obviously, a data reduction must be performed to determine which of the samples are
displayed for each sweep point. This is the trace detector's task.
The trace detector can analyze the measured data using various methods:
Trace Configuration

Table of Contents

Other manuals for R&S FSV3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSV3000 Series and is the answer not in the manual?

R&S FSV3000 Series Specifications

General IconGeneral
TOI (Third-Order Intercept)> +15 dBm
Resolution Bandwidth1 Hz to 1 MHz
Display12.1-inch touchscreen
Analysis Bandwidth40 MHz
Video Bandwidth (VBW)1 Hz to 1 MHz
Amplitude accuracy±0.5 dB
InterfaceUSB
Dynamic Range> 100 dB

Related product manuals