General Instrument Setup
R&S
®
FSVA3000/ R&S
®
FSV3000
620User Manual 1178.8520.02 ─ 01
If the number of error messages exceeds the capacity of the error buffer, "Message
Buffer Overflow" is displayed. To clear the message buffer use the "Clear All Mes-
sages" button.
The following information is available:
No device-specific error code
Message brief description of the message
Component hardware messages: name of the affected module
software messages: name of the affected software
Date/Time date and time of the occurrence of the message
Remote command:
SYSTem:ERRor:LIST? on page 1132
12.8.4 Firmware Updates
Access: [Setup] > "System Configuration" > "Firmware Update"
During instrument start, the installed hardware is checked against the current firmware
version to ensure the hardware is supported. If not, an error message is displayed
( "Wrong Firmware Version" ) and you are asked to update the firmware. Until the firm-
ware version is updated, self-alignment fails. To see which components are not suppor-
ted, see the System Messages.
The firmware on your R&S FSV/A may also need to be updated in order to enable
additional new features or if reasons for improvement come up. Ask your sales repre-
System Configuration Settings

Table of Contents