Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
167User Manual 1178.8520.02 ─ 01
Remote command:
[SENSe:]POWer:ACHannel:NAME:ACHannel on page 755
[SENSe:]POWer:ACHannel:NAME:ALTernate<ch> on page 755
[SENSe:]POWer:ACHannel:NAME:CHANnel<ch> on page 756
7.3.5 MSR ACLR Configuration
Access: "Overview" > "Select Measurement" > "Channel Power ACLR" > "CP / ACLR
Standard" > "Standard" : "Multi-Standard Radio" > "CP / ACLR Config"
ACLR measurements can also be performed on input containing multiple signals for
different communication standards. A new measurement standard is provided that
allows you to define multiple discontiguous transmit channels at specified frequencies,
independent from the selected center frequency. If the "Multi-Standard Radio" standard
is selected (see " Standard " on page 159), the "ACLR Setup" dialog box is replaced by
the "MSR ACLR Setup" dialog box.
For more information, see Chapter 7.3.3.4, "Measurement on Multi-Standard Radio
(MSR) Signals", on page 153.
The remote commands required to perform these tasks are described in Chap-
ter 14.5.3, "Measuring the Channel Power and ACLR", on page 752.
● General MSR ACLR Measurement Settings.........................................................167
● MSR Sub Block and Tx Channel Definition...........................................................172
● MSR Adjacent Channel Setup.............................................................................. 175
● MSR Gap Channel Setup......................................................................................178
● MSR Channel Names........................................................................................... 180
7.3.5.1 General MSR ACLR Measurement Settings
Access: "Overview" > "Select Measurement" > "Channel Power ACLR" > "CP / ACLR
Standard" > "Standard" : "Multi-Standard Radio" > "CP / ACLR Config" > "MSR Gen-
eral Settings" tab
Channel Power and Adjacent-Channel Power (ACLR) Measurement