Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1108User Manual 1178.8520.02 ─ 01
[SENSe:]CORRection:FRESponse:USER:IQ:DATA:FREQuency?
[SENSe:]CORRection:FRESponse:USER:IQ:DATA:MAGNitude?
[SENSe:]CORRection:FRESponse:USER:IQ:DATA:PHASe?
Queries the trace values for the combined user correction files (.snp+.fres) in IQ
mode.
Example:
SENS:CORR:FRES:USER:IQ:DATA:PHAS?
Usage: Query only
Manual operation: See "IQ Mode" on page 612
[SENSe:]CORRection:FRESponse:BASeband:USER:LOAD <FilePath>
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:LOAD <FilePath>
[SENSe:]CORRection:FRESponse:USER:LOAD <FilePath>
Loads a stored user-defined frequency response correction scenario.
Setting parameters:
<FilePath> string
Example:
SENS:CORR:FRES:USER:LOAD 'FRes1'
Usage: Setting only
Manual operation: See "Load Settings" on page 608
[SENSe:]CORRection:FRESponse:BASeband:USER:PRESet
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:PRESet
[SENSe:]CORRection:FRESponse:USER:PRESet
Restores the default frequency response correction settings (containing only files spe-
cific to the R&S FSV/A itself). Frequency response correction using .fres files is
deactivated.
Example:
SENS:CORR:FRES:USER:PRES
Usage: Event
Manual operation: See "Clear Settings" on page 609
[SENSe:]CORRection:FRESponse:USER:SCOPe <Scope>
Determines whether the frequency response correction settings are applied to all
active measurement channels, or only the currently selected channel.
Parameters:
<Scope> CHANnel | ALL
CHANnel
The frequency response correction settings are applied to the
currently selected channel only.
To select a channel, use INSTrument[:SELect].
For a list of available channels, use INSTrument:LIST?.
Configuring the R&S FSV/A