Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
988User Manual 1178.8520.02 ─ 01
If the marker has been used as a delta marker, the command turns it into a normal
marker.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Position> Numeric value that defines the marker position on the x-axis.
The unit depends on the result display.
Range: The range depends on the current x-axis range.
Default unit: Hz
Example:
CALC:MARK2:X 1.7MHz
Positions marker 2 to frequency 1.7 MHz.
Manual operation: See " Marker 1 / Marker 2 / Marker 3 / Marker 4 " on page 298
See " Marker 1 / Marker 2 / Marker 3 " on page 303
See " Marker Table " on page 413
See " Marker Peak List " on page 413
See "Marker Position X-value " on page 431
14.8.3.2 General Marker Settings
The following commands control general marker functionality.
Remote commands exclusive to general marker functionality
DISPlay[:WINDow<n>]:MTABle....................................................................................... 988
DISPlay[:WINDow<n>]:MINFo[:STATe].............................................................................989
CALCulate<n>:MARKer<m>:X:SSIZe.............................................................................. 989
DISPlay[:WINDow<n>]:MTABle <DisplayMode>
This command turns the marker table on and off.
Suffix:
<n>
.
irrelevant
Parameters:
<DisplayMode> ON | 1
Turns on the marker table.
OFF | 0
Turns off the marker table.
AUTO
Turns on the marker table if 3 or more markers are active.
*RST: AUTO
Example:
DISP:MTAB ON
Activates the marker table.
Manual operation: See " Marker Table Display " on page 433
Analyzing Measurements (Basics)