Common Analysis and Display Functions
R&S
®
FSVA3000/ R&S
®
FSV3000
431User Manual 1178.8520.02 ─ 01
Remote command:
CALCulate<n>:MARKer<m>[:STATe] on page 987
CALCulate<n>:DELTamarker<m>[:STATe] on page 985
Marker Position X-value
Defines the position (x-value) of the marker in the diagram. For normal markers, the
absolute position is indicated. For delta markers, the position relative to the reference
marker is provided.
Remote command:
CALCulate<n>:MARKer<m>:X on page 987
CALCulate<n>:DELTamarker<m>:X on page 985
Frame (Spectrogram only)
Spectrogram frame the marker is assigned to.
Remote command:
CALCulate<n>:MARKer<m>:SPECtrogram:FRAMe on page 1002
CALCulate<n>:DELTamarker<m>:SPECtrogram:FRAMe on page 1006
Marker Type
Toggles the marker type.
The type for marker 1 is always "Normal" , the type for delta marker 1 is always
"Delta" . These types cannot be changed.
Note: If normal marker 1 is the active marker, switching the "Mkr Type" activates an
additional delta marker 1. For any other marker, switching the marker type does not
activate an additional marker, it only switches the type of the selected marker.
"Normal"
A normal marker indicates the absolute value at the defined position
in the diagram.
"Delta"
A delta marker defines the value of the marker relative to the speci-
fied reference marker (marker 1 by default).
Remote command:
CALCulate<n>:MARKer<m>[:STATe] on page 987
CALCulate<n>:DELTamarker<m>[:STATe] on page 985
Reference Marker
Defines a marker as the reference marker which is used to determine relative analysis
results (delta marker values).
Remote command:
CALCulate<n>:DELTamarker<m>:MREFerence on page 984
Linking to Another Marker
Links the current marker to the marker selected from the list of active markers. If the x-
axis value of the initial marker is changed, the linked marker follows to the same posi-
tion on the x-axis. Linking is off by default.
Using this function you can set two markers on different traces to measure the differ-
ence (e.g. between a max hold trace and a min hold trace or between a measurement
and a reference trace).
Marker Usage