Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
834User Manual 1178.8520.02 ─ 01
Example:
CALC:PSE:AUTO OFF
Deactivates the list evaluation.
CALCulate<n>:PSEarch:MARGin <Threshold>
CALCulate<n>:PEAKsearch:MARGin <Margin>
This command defines the threshold of the list evaluation.
Suffix:
<n>
.
Window
Parameters:
<Margin> Range: -200 to 200
Default unit: dB
Example:
CALC:PSE:MARG 100
Sets the threshold to 100 dB.
CALCulate<n>:PSEarch:PSHow
<State>
CALCulate<n>:PEAKsearch:PSHow <State>
This command turns the peak labels in the diagram on and off.
Peak labels are blue squares.
Suffix:
<n>
.
Window
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:PSE:PSH ON
Marks all peaks with blue squares.
CALCulate<n>:PSEarch:SUBRanges <NumberPeaks>
CALCulate<n>:PEAKsearch:SUBRanges <NumberPeaks>
This command defines the number of peaks included in the peak list.
After this number of peaks has been found, the R&S FSV/A stops the peak search and
continues the search in the next measurement range.
Suffix:
<n>
.
Window
Parameters:
<NumberPeaks> Range: 1 to 50
*RST: 25
Example:
CALC:PSE:SUBR 10
Sets 10 peaks per range to be stored in the list.
Manual operation: See " Peaks per Range " on page 261