Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
874User Manual 1178.8520.02 ─ 01
<TriggerOffset> Defines the trigger delay.
Range: 0 s to 30 s
*RST: 0 s
Default unit: s
<Detector> Selects the detector and therefore the way the measurement is
evaluated.
MEAN
Calculates the RMS pulse power.
PEAK
Calculates the peak pulse power.
<OfPulses> Defines the number of pulses included in the measurement.
Range: 1 to 32001
Return values:
<PowerLevel> Pulse power level.
The result is a list of floating point values separated by commas.
The unit is dBm.
[SENSe:]MPOWer:RESult:MIN?
This command queries the lowest pulse power that has been measured during a pulse
power measurement.
If there are no results, the command returns an error.
Return values:
<PulsePower> Lowest power level of the pulse power measurement.
The unit is dBm.
Usage: Query only
14.5.14.2 Example: Performing a Pulse Power Measurement
The following example shows a pulse power measurement.
-----Measurement with synchronization via service request-----
*ESE 1
*SRE 32
// Configures the status reporting system to produce a service request.
MPOW:FTYP NORM
//Selects a Gaussian filter for the measurement.
MPOW 935.2MHZ,1MHZ,434us,VID,50,5us,MEAN,20;
*OPC
//Configures and initiates a measurement on 20 pulses with synchronization to the end.
//Analyzer produces a service request
//On service request:
MPOW:RES?
//Returns the results of the measurements (20 power levels).
MPOW:RES:MIN?
//Returns the lowest of the 20 power level that have been measured.
Configuring and Performing Measurements