Network and Remote Operation
R&S
®
FSVA3000/ R&S
®
FSV3000
703User Manual 1178.8520.02 ─ 01
Language.................................................................................................................... 703
IF Gain........................................................................................................................ 704
Sweep Repeat.............................................................................................................704
Coupling......................................................................................................................704
Wideband....................................................................................................................704
FSU/FSQ Preamplifier................................................................................................ 705
HP Additional.............................................................................................................. 705
Revision String............................................................................................................705
Resetting the Factory Revision................................................................................... 705
Language
Defines the system language used to control the instrument.
For details on the available GPIB languages, see Chapter 14.13.2, "Reference: GPIB
Commands of Emulated HP Models", on page 1160.
Note: Emulating previous R&S signal and spectrum analyzers. This function is also
used to emulate previous R&S signal and spectrum analyzers.
As a rule, the R&S FSV/A supports most commands from previous R&S signal and
spectrum analyzers such as the FSQ, FSP, FSU, or FSV. However, the default values,
in particular the number of sweep points or particular bandwidths, may vary. Therefore,
the R&S FSV/A can emulate these other devices, including their default values, in
order to repeat previous measurements or support existing control applications as in
legacy systems.
For R&S FSP/FSQ/FSU emulation, HP commands are not automatically also allowed.
In this case, set SYSTem:HPADditional to ON.
Note: For PSA89600 emulation, the option is indicated as "B7J" for the *OPT? query
("B7J, 140" or "B7J, 122" if Wideband is activated, see SYSTem:PSA:WIDeband
on page 1159).
Network and Remote Control Settings