Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1028User Manual 1178.8520.02 ─ 01
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 1.
CALC:MARK:COUN ON
Switches on the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value.
Manual operation: See " Signal Count Marker State " on page 444
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
This command queries the frequency at the marker position.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 746.
Before you can use the command, you have to turn on the frequency counter.
Suffix:
<n>
.
Window
<m> Marker
Return values:
<Frequency> Frequency at the marker position.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 2.
CALC:MARK:COUN ON
Activates the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value of marker 1.
Usage: Query only
Manual operation: See " Signal Count Marker State " on page 444
CALCulate<n>:MARKer<m>:COUNt:RESolution <Resolution>
This command defines the resolution of the frequency counter.
Suffix:
<n>
.
Window
<m> Marker
Analyzing Measurements (Basics)