Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1092User Manual 1178.8520.02 ─ 01
ON | 1
Switches the reference on
Example: //Output reference signal of 100 MHz.
ROSC:O100 ON
Manual operation: See "Reference Frequency Output" on page 616
SOURce:EXTernal:ROSCillator:EXTernal:FREQuency <Frequency>
This command defines the frequency of the external reference oscillator.
If the external reference oscillator is selected, the reference signal must be connected
to the rear panel of the instrument.
Example:
ROSC:EXT:FREQ 13MHZ
Sets the frequency to 13 MHz.
SOUR:EXT:ROSC:EXT:FREQ 13MHZ
Manual operation: See "Reference Frequency Input" on page 615
[SENSe:]ROSCillator:SOURce
<Source>
This command selects the reference oscillator.
If you want to select the external reference, it must be connected to the R&S FSV/A.
Parameters:
<Source> INTernal
The internal reference is used (10 MHz)
EXTernal | EXTernal1 | EXT1
The external reference from the "REF INPUT 10 MHZ" connec-
tor is used; if none is available, an error flag is displayed in the
status bar
EAUTo
The external reference is used as long as it is available, then the
instrument switches to the internal reference
Example:
ROSC:SOUR EXT
Manual operation: See "Reference Frequency Input" on page 615
See "Behavior in case of missing external reference"
on page 616
[SENSe:]ROSCillator:SOURce:EAUTo?
This command queries the current reference type in case you have activated an auto-
matic switch to the internal reference if the external reference is missing.
Return values:
<Reference> INT | EXT
INT
internal reference
Configuring the R&S FSV/A