Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
158User Manual 1178.8520.02 ─ 01
Both Channel Power (CP) and Adjacent-Channel Power (ACLR) measurements are
available.
If the "Multi-Standard Radio" standard is selected (see " Standard " on page 159), the
"ACLR Setup" dialog box is replaced by the "MSR ACLR Setup" dialog box. See Chap-
ter 7.3.5, "MSR ACLR Configuration", on page 167 for a description of these settings.
Auto measurement
Use the Auto All function to determine the required measurement parameters automat-
ically.
The following settings are determined:
Center frequency
Span: determined signal bandwidth plus 40 %
TX and adjacent channels
ACLR standard according to estimated signal bandwidth
For pulsed signals: gating parameters
See also Chapter 7.1, "Auto Measurements", on page 121.