EasyManua.ls Logo

R&S FSV3000 Series - Page 367

R&S FSV3000 Series
1268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
367User Manual 1178.8520.02 ─ 01
The reference level determines the amplitude represented by the topmost grid line in
the display. When you change the reference level, the measurement is not restarted;
the results are merely shifted in the display. Only if the reference level changes due to
a coupled RF attenuation (see " Attenuation Mode / Value " on page 371), the mea-
surement is restarted.
In general, the R&S FSV/A measures the signal voltage at the RF input. The level dis-
play is calibrated in RMS values of an unmodulated sine wave signal. In the default
state, the level is displayed at a power of 1 mW (= dBm). Via the known input impe-
dance, conversion to other units is possible.
See " Impedance " on page 313.
Reference level offset
If the signal is attenuated or amplified before it is fed into the R&S FSV/A, you can
define an (arithmetic) offset to the reference level so the application shows correct
power results. All displayed power level results are shifted by this value, and the scal-
ing of the y-axis is changed accordingly.
To determine the required offset, consider the external attenuation or gain applied to
the input signal. For attenuation, define a positive offset so the R&S FSV/A increases
the displayed power values.
If an external gain is applied, define a negative offset so the R&S FSV/A decreases the
displayed power values.
Note, however, that the internal reference level (used to adjust the hardware settings to
the expected signal optimally) ignores any "Reference Level Offset" . Thus, it is impor-
tant to keep in mind the actual power level the R&S FSV/A must handle, and not to rely
on the displayed reference level.
internal reference level = displayed reference level - offset
Example
1. The initial reference level is 2 dBm with no offset.
Both the displayed reference level and the internal reference level are 2 dBm.
2. An offset of 3 dB is defined.
The displayed reference level is adjusted to 5 dBm.
The internal reference level remains at 2 dBm.
(5 dBm (displayed ref level) - 3 dB (offset) = 2 dBm)
3. Now the user decreases the reference level to 1 dBm.
The displayed reference level is adjusted to 1 dBm.
The internal reference level is adjusted to:
1 dBm (displayed ref level) - 3 dB (offset) = -2 dBm.
Amplitude and Vertical Axis Configuration

Table of Contents

Other manuals for R&S FSV3000 Series

Related product manuals