Common Analysis and Display Functions
R&S
®
FSVA3000/ R&S
®
FSV3000
484User Manual 1178.8520.02 ─ 01
How to export a limit line
Limit line configurations can be stored to an ASCII file for evaluation in other programs
or to be imported later for other measurements.
1. In the "Line Config" dialog box, select the limit line.
2. Select the "New" or "Edit" button.
3. Define the limit line as described in "How to configure a new limit line"
on page 482.
4. Select "Export" to save the configuration to a file.
You are asked whether you would like to save the configuration internally on the
R&S FSV/A first.
5. Select a file name and location for the limit line.
6. Select the decimal separator to be used in the file.
7. Select "Save" .
The limit line is stored to a file with the specified name and the extension .CSV.
For details on the file format see Chapter 9.4.2.4, "Reference: Limit Line File For-
mat", on page 485.
How to import a limit line
Limit line configurations that are stored in an ASCII file and contain a minimum of
required data can be imported to the R&S FSV/A.
For details on the required file format see Chapter 9.4.2.4, "Reference: Limit Line File
Format", on page 485.
1. In the "Line Config" dialog box, select the limit line.
2. Select the "New" or "Edit" button.
3. Select "Import" to load a limit line from a file.
You are asked whether you would like to save the current configuration on the
R&S FSV/A first.
4. Select the file name of the limit line.
5. Select the decimal separator that was used in the file.
6. Select "Select" .
The limit line is loaded from the specified file and displayed in the "Edit Limit Line"
dialog box.
7. Activate the limit line as described in "How to activate and deactivate a limit check"
on page 481.