General Instrument Setup
R&S
®
FSVA3000/ R&S
®
FSV3000
587User Manual 1178.8520.02 ─ 01
[Time]: =[Hour]-[Minute]-[Second]
Example:
The definition ScreenShot_[Date]_[Time] creates file names such as
ScreenShot_2018-12-14_12-49-51.PNG,
ScreenShot_2018-12-14_12-50-01.PNG.
Format ← Store Screenshot ← Choose Action
Defines the image file format
Stop Cont Sweep ← Choose Action
Stops a continuous sweep.
No further parameters required.
Deactivate Rule ← Choose Action
Deactivates the rule after the specified number of trigger events occurred; useful to
perform a single action only.
Auto Level ← Choose Action
Automatically optimizes the level settings according to the current measurement;
If several "Auto Level" actions are defined for the same rule or for multiple rules, only
one "Auto Level" action is performed.
Start Command ← Choose Action
Executes the specified command, for example a user-defined script or an external
application. The command is executed asynchronously to the measurement, that is:
the measurement continues without waiting for the command to finish execution.
Type ← Start Command ← Choose Action
Defines the type of command to be executed, which determines the application
required to execute the command. For example, a SCPI command (*.inp) is execu-
ted in the R&S IECWIN tool.
Command (Browse) ← Start Command ← Choose Action
Defines the path and file name of the command to be executed; invalid (not available)
command names are indicated by red font.
Event-based Actions