General Instrument Setup
R&S
®
FSVA3000/ R&S
®
FSV3000
603User Manual 1178.8520.02 ─ 01
How to move the transducer line vertically or horizontally
A configured transducer line can easily be moved vertically or horizontally. Thus, a new
transducer line can be easily generated based upon an existing transducer line which
has been shifted.
1. In the "Line Config" dialog box, select the transducer line.
2. Select the "Edit" button.
3. In the "Edit Transducer Line" dialog box, select the "Shift x" or "Shift y" button and
define the shift width.
4. Save the shifted data points by selecting the "Save" button.
If activated, the trace is recalculated after the next sweep.
How to export a transducer factor
Transducer factor configurations can be stored to an ASCII file for evaluation in other
programs or to be imported later for other measurements.
1. In the "Edit Transducer" dialog box, select the transducer factor.
2. Select the "New" or "Edit" button.
3. Define the transducer factor as described in "How to configure a new transducer
line" on page 602.
4. Select "Export" to save the configuration to a file.
You are asked whether you would like to save the configuration internally on the
R&S FSV/A first.
5. Select a file name and location for the transducer factor.
6. Select the decimal separator to be used in the file.
7. Select "Save" .
The transducer factor is stored to a file with the specified name and the exten-
sion .CSV.
For details on the file format see Chapter 12.5.3, "Reference: Transducer Factor
File Format", on page 599.
How to import a transducer factor
Transducer factor configurations that are stored in an ASCII file and contain a minimum
of required data can be imported to the R&S FSV/A.
For details on the required file format see Chapter 12.5.3, "Reference: Transducer Fac-
tor File Format", on page 599.
1. In the "Edit Transducer" dialog box, select the transducer factor.
2. Select the "New" or "Edit" button.
3. Select "Import" to load a transducer factor from a file.