Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
778User Manual 1178.8520.02 ─ 01
For more information see Chapter 7.3.3.4, "Measurement on Multi-Standard Radio
(MSR) Signals", on page 153.
Parameters:
<State> ON | OFF | 1 | 0
ON | 1
The gap channels are displayed and channel power results are
calculated and displayed in the Result Summary.
OFF | 0
The gap channels are not displayed in the diagram and channel
power results are not calculated nor displayed in the Result
Summary.
*RST: 1
Manual operation: See " Activate Gaps " on page 179
[SENSe:]POWer:ACHannel:BWIDth:GAP<gap> <Bandwidth>
[SENSe:]POWer:ACHannel:BANDwidth:GAP<gap> <Bandwidth>
This command defines the bandwidth of the specified MSR gap (CACLR) channel.
This command is for MSR signals only (see CALCulate<n>:MARKer<m>:
FUNCtion:POWer<sb>:PRESet on page 752).
Suffix:
<gap>
.
1..n
Gap (CACLR) channel number
Parameters:
<Bandwidth> numeric value in Hz
*RST: 3.84 MHz
Default unit: Hz
Manual operation: See " Gap Channel Bandwidths " on page 180
[SENSe:]POWer:ACHannel:FILTer:ALPHa:GAP<gap> <Alpha>
This command defines the roll-off factor for the specified gap (CACLR) channel's
weighting filter.
Suffix:
<gap>
.
1..n
Gap (CACLR) channel number
Parameters:
<Alpha> Roll-off factor
Range: 0 to 1
*RST: 0.22
Manual operation: See " Weighting Filters " on page 180