Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
846User Manual 1178.8520.02 ─ 01
Usage: Query only
CALCulate<n>:STATistics:RESult<res>? <ResultType>
This command queries the results of a CCDF or ADP measurement for a specific
trace.
Suffix:
<n>
.
irrelevant
<res> Trace
Query parameters:
<ResultType> MEAN
Average (=RMS) power in dBm measured during the measure-
ment time.
PEAK
Peak power in dBm measured during the measurement time.
CFACtor
Determined crest factor (= ratio of peak power to average
power) in dB.
ALL
Results of all three measurements mentioned before, separated
by commas: <mean power>,<peak power>,<crest factor>
Example:
CALC:STAT:RES2? ALL
Reads out the three measurement results of trace 2. Example of
answer string: 5.56,19.25,13.69 i.e. mean power: 5.56 dBm,
peak power 19.25 dBm, crest factor 13.69 dB
Usage: Query only
14.5.8.7 Programming Example: Measuring Statistics
This example demonstrates how to determine statistical values for a measurement in a
remote environment using the gated statistics example described in Chapter 7.8.4,
"APD and CCDF Basics - Gated Triggering", on page 267.