EasyManua.ls Logo

R&S FSV3000 Series - Page 851

R&S FSV3000 Series
1268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
851User Manual 1178.8520.02 ─ 01
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak[:STATe] <State>
This command turns the evaluation to determine the positive peak time domain power
on and off.
The R&S FSV/A performs the measurement on the trace marker 1 is positioned on.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See " Results " on page 282
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS[:STATe]
<State>
This command turns the evaluation to determine the RMS time domain power on and
off.
The R&S FSV/A performs the measurement on the trace marker 1 is positioned on.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See " Results " on page 282
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation[:STATe] <State>
This command turns the evaluation to determine the standard deviation of the time
domain power on and off.
The R&S FSV/A performs the measurement on the trace marker 1 is positioned on.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
14.5.9.2 Performing a Time Domain Power Measurement
The following commands are required to perform a Time Domain Power measurement:
Configuring and Performing Measurements

Table of Contents