Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
854User Manual 1178.8520.02 ─ 01
See also INITiate<n>:CONTinuous on page 746.
Suffix:
<n>
.
Window
<m> Marker
Return values:
<PeakPower> Peak power of the signal during the measurement time.
Usage: Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak:RESult?
This command queries the positive peak time domain power.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 746.
Suffix:
<n>
.
Window
<m> Marker
Return values:
<PeakPower> Peak power of the signal during the measurement time.
Usage: Query only
Manual operation: See " Results " on page 282
Measuring the RMS Power
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:AVERage:RESult?
This command queries the average RMS of the time domain power. The query is only
possible if averaging has been activated previously using CALCulate<n>:
MARKer<m>:FUNCtion:SUMMary:AVERage on page 849.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 746.
Suffix:
<n>
.
Window
<m> Marker
Return values:
<RMSPower> RMS power of the signal during the measurement time.
Usage: Query only