Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
945User Manual 1178.8520.02 ─ 01
Suffix:
<p>
.
Power sensor index
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
PMET2:DCYC:STAT ON
Manual operation: See " Duty Cycle " on page 321
[SENSe:]PMETer<p>:DCYCle:VALue <Percentage>
This command defines the duty cycle for the correction of pulse signals.
The power sensor uses the duty cycle in combination with the mean power to calculate
the power of the pulse.
Suffix:
<p>
.
Power sensor
Parameters:
<Percentage> Range: 0.001 to 99.999
*RST: 99.999
Default unit: %
Example:
PMET2:DCYC:STAT ON
Activates the duty cycle correction.
PMET2:DCYC:VAL 0.5
Sets the correction value to 0.5%.
Manual operation: See " Duty Cycle " on page 321
[SENSe:]PMETer<p>:FREQuency <Frequency>
This command defines the frequency of the power sensor.
Suffix:
<p>
.
Power sensor index
Parameters:
<Frequency> The available value range is specified in the data sheet of the
power sensor in use.
*RST: 50 MHz
Default unit: HZ
Example:
PMET2:FREQ 1GHZ
Sets the frequency of the power sensor to 1 GHz.
Manual operation: See " Frequency Manual " on page 320
Setting Basic Measurement Parameters