EasyManua.ls Logo

R&S SMB100A - Page 359

R&S SMB100A
528 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMB100A
344Operating Manual 1407.0806.32 ─ 21
Parameters:
<Mode> CW | FIXed | SWEep | LIST
CW|FIXed
Sets the fixed frequency mode.
CW and FIXed are synonyms. The instrument operates at a
defined frequency, set with command [:SOURce<hw>]:
FREQuency[:CW|FIXed]..
SWEep
Sets the sweep mode. The instrument processes the frequency
settings in defined sweep steps. To determine the corresponding
frequency values, use the commands [:SOURce<hw>]:
FREQuency:STARt and [:SOURce<hw>]:FREQuency:STOP,
or [:SOURce<hw>]:FREQuency:CENTer and [:
SOURce<hw>]:FREQuency:SPAN and [:SOURce<hw>]:
FREQuency:MANual.
LIST
Sets the list mode. The instrument processes the frequency and
level settings by means of values loaded from a list.
To configure the list mode settings use the commands of the
SOURce:LIST Subsystem
*RST:
CW
Example:
FREQ:MODE SWE
sets the SWEep mode.
Manual operation: See "State - Frequency Sweep" on page 176
[:SOURce<hw>]:FREQuency:MULTiplier <Multiplier>
Sets the value for the multiplication factor of a subsequent downstream instrument.
Parameters:
<Multiplier> float
Range: 1 to dynamic
Increment: 0.001
*RST: 1
Example:
FREQ:MULT 1
sets the multiplication factor to 1.
Manual operation: See "Multiplier" on page 139
[:SOURce<hw>]:FREQuency:OFFSet <Offset>
Sets the frequency offset of a downstream instrument, for example a mixer.
If you have specified an OFFSet and / or a MULTiplier factor, the actual frequency
at the RF output does not change, but rather the value queried with [:SOUR]:FREQ?,
according to the following formula:
f
FREQ
= f
RFout
* f
MULTiplier
+ f
OFFSet
SOURce Subsystem

Table of Contents

Other manuals for R&S SMB100A