Fernsteuerbefehle
R&S
®
RTB2000
297Bedienhandbuch 1333.1611.03 ─ 05
Parameter:
<Time1> Center value, maximum value or minimum value depending on
the defined range type.
Bereich: 20E-9 bis 6.87194685440
Inkrement: Depends on the <Time1> value
*RST: 20E-9
TRIGger:A:WIDTh:DELTa <Delta>
Defines a variation range around the width value specified using TRIGger:A:WIDTh:
WIDTh.
Parameter:
<Delta> Variation ±Δt
Bereich: Minimum is 0. Maximum depends on the defined
pulse width (TRIG:A:WIDTH:WITDH).
15.5.4 Video/TV-Trigger
TRIGger:A:TV:STANdard................................................................................................ 297
TRIGger:A:TV:POLarity.................................................................................................. 297
TRIGger:A:TV:FIELd...................................................................................................... 298
TRIGger:A:TV:LINE........................................................................................................298
TRIGger:A:TV:STANdard <Standard>
Selects the color television standard.
Parameter:
<Standard> PAL | NTSC | SECam | PALM | I576 | P720 | P1080 | I1080
PALM = PAL-M
I576 = SDTV 576i (PAL and SECAM)
P720 | P1080 = HDTV 720/1080p (progressive scanning)
I1080 = HDTV 1080i (interlaced scanning)
*RST: PAL
TRIGger:A:TV:POLarity <Polarity>
Selects the polarity of the signal. Note that the sync pulse has the opposite polarity.
The edges of the sync pulses are used for triggering,
See also: "Signal" auf Seite 67