Fernsteuerbefehle
R&S
®
RTB2000
334Bedienhandbuch 1333.1611.03 ─ 05
MEASurement<m>:AOFF
Stops the quick measurement.
Verwendung: Ereignis
MEASurement<m>:ARESult?
Returns the results of the quick measurement.
Rückgabewerte:
<Data> List of values
Quick measurement results are listed in the following order:
PEAK (Vpp), UPE (Vp+), LPE (Vp-), CYCR (RMS-Cyc), CYCM
(MeanCyc), PER (T), FREQ (f), RTIM (tr), FTIM (tf).
Verwendung: Nur Abfrage
MEASurement<m>:ALL[:STATe]
Starts or stops the quick measurement and sets the status bit.
Suffix:
<m>
.
1..4
The suffix is irrelevant.
Parameter:
<State> ON | OFF
*RST: OFF
15.7.2 Automatische Messungen
15.7.2.1 Messeinstellungen
MEASurement<m>[:ENABle]...........................................................................................334
MEASurement<m>:MAIN................................................................................................335
MEASurement<m>:SOURce...........................................................................................337
MEASurement<m>:DELay:SLOPe...................................................................................338
MEASurement<m>:STATistics[:ENABle]...........................................................................338
MEASurement<m>:STATistics:RESet...............................................................................339
MEASurement<m>[:ENABle] <State>
Activates or deactivates the selected measurement.
Suffix:
<m>
.
1..4
Selects the measurement place.