Fernsteuerbefehle
R&S
®
RTB2000
492Bedienhandbuch 1333.1611.03 ─ 05
Parameter:
<SampleTime> Numeric value
Bereich:  2.000E-08  bis  4.200E+01
*RST:  2.000E-08
Std.-einheit: s
PGENerator:PATTern:PERiod <PatternPeriod>
Defines the period of the pattern generator function.
Parameter:
<PatternPeriod> Numeric value (Period = Pattern length * Bit time)
Bereich:  MIN 1 Sample * 20ns = 20ns to MAX 2048 Samp-
les * 42s = 10416s (approx. 2.89h)
*RST:  2.000E-06
Std.-einheit: s
PGENerator:PATTern:FREQuency <PatternFrequency>
Defines the frequency (period) value of the pattern generator function.
Parameter:
<PatternFrequency> Numeric value
Std.-einheit: Hz
PGENerator:PATTern:ITIMe <IdleTime>
Defines the idle time of the pattern generator function. The idle time can be only defi-
ned with activated BURST function.
Parameter:
<IdleTime> Numeric value
Bereich:  2.000E-08  bis  4.200000000000E+01
*RST:  2.5000000000E-01
Std.-einheit: s
PGENerator:PATTern:BURSt:STATe <BurstState>
Turns the BURST function on or off.
Parameter:
<BurstState> ON | OFF
*RST:  OFF
PGENerator:PATTern:BURSt:NCYCle <PatternCycles>
Defines the BURST pattern cycles. The cycles can be only defined with activated
BURST function.
Signalerzeugung (Option R&S
 RTB-B6)