20077/021 US XTRA Operator’s Manual
6. Using the pump handle, manually turn the pump rotor clockwise until all the remaining fluid in
the centrifuge bowl has been emptied into the Empty bag.
7. Remove the pump loop tubing from the pump rotor. To do so, turn the pump rotor counter
clockwise while pulling the pump loop tubing free.
8. Fold the pump handle back to its closed position, lining up the tab on the bottom of the handle
with the groove in the pump rotor.
9. Dismount the bowl and the rest of the disposable set.
10. Close the centrifuge lid.
Note: In case the power loss occurs before washing of the blood is complete, the operator may
decide to manually return the bowl content to the reservoir. In this case, the same steps as above can
be followed with differences to these two steps:
4. Unclamp the Fill line so that fluid can flow from the bowl to the Reservoir.
5. Using the pump handle, manually turn the pump rotor clockwise until all the remaining fluid in
the centrifuge bowl has been emptied into the reservoir.