Preoperative Sequestration (PPP and PRP)
10-18
20077/021 US XTRA Operator’s Manual
PPP Bag Full
The PPP Bag Full warning is triggered during the Fill or Spill phases if the volume transferred to the
PPP bag exceeds the pre-set safety level. Also refer to “PPP bag full. Empty or replace it.” on page
14-18.
When triggered, the acoustic sequence is sounded (if enabled), the message “PPP bag full. Empty or
replace it.” is displayed in the message area of the current screen, the PPP/PRP displet is highlighted
with a red outline, and the processing cycle is paused.
To proceed, empty or replace the PPP bag, then touch the Play button to resume.
Note: The volume counter will be reset to zero.
PRP Bag Full
The PRP Bag Full warning is triggered during the Fill or Spill phases if the volume transferred to the
PRP bag exceeds the pre-set safety level. Also refer to “PRP bag full. Empty or replace it.” on page
14-18.
When triggered, the acoustic sequence is sounded (if enabled), the message “PRP bag full. Empty or
replace it.” is displayed in the message area of the current screen, the PPP/PRP displet is highlighted
with a red outline, and the processing cycle is paused.
To proceed, empty or replace the PRP bag, then touch the Play button to resume.
Note: The volume counter will be reset to zero.
RBC Bag Full
The RBC Bag Full warning is triggered during the Empty phase if the volume transferred to the RBC
bag exceeds the pre-set safety level. Also refer to “RBC bag full. Empty or replace it.” on page 14-19.
When triggered, the acoustic sequence is sounded (if enabled), the message RBC bag full. Empty or
replace it. is displayed in the message area of the current screen, the RBC displet is highlighted with a
red outline, and the processing cycle is paused.
To proceed, empty or replace the RBC bag, then touch the Play button to resume.
Note: The volume counter will be reset to zero.