System Description
X
TRA Operators Manual 20077/021 US 3-25
is strongly recommended that a Microaggregate Blood Filter (20 to 40 micron) be used on the
reinfusion line. These two items are not
included with the XTRA Bowl Set.
Procedure Set
The Procedure set is equivalent to the Collection Set CARDIO and the Bowl Set (see Figure 3-35). The
Procedure set contains all the components needed for blood collection and processing:
Collection Set CARDIO (TOP or BOTTOM)
◊ Blood Collection Reservoir (TOP or BOTTOM)
◊ Aspiration Line
◊ Vacuum Line
Bowl Set (55 ml, 125 ml, 175 ml, or 225 ml)
Nine configurations of the X
TRA Procedure set are available:
Procedure set with TOP reservoir and 55 ml bowl
Procedure set with TOP reservoir and 125 ml bowl
Procedure set with TOP reservoir and 175 ml bowl
Procedure set with TOP reservoir and 225 ml bowl
Procedure set with BOTTOM reservoir and 55 ml bowl
Procedure set with BOTTOM reservoir and 125 ml bowl
Procedure set with BOTTOM reservoir and 175 ml bowl
Procedure set with BOTTOM reservoir and 225 ml bowl
Procedure set with BOTTOM reservoir 120 μm and 225 ml bowl
WARNING
Do not reinfuse the patient’s blood from the primary RBC bag when
it is connected to the X
TRA autologous transfusion circuit. Reinfusion
from the primary reinfusion bag connected to the circuit could lead
to air embolism.
WARNING
To minimize the complications of particulate matter infusion and the
risk of air embolism, use of an in-line microaggregate filter on the
patient reinfusion line is STRONGLY RECOMMENDED.
WARNING
To reduce risk of air embolism, remove all air from the primary
reinfusion bag before handing the bag over for reinfusion.