Preoperative Sequestration (PPP and PRP)
X
TRA Operators Manual 20077/021 US 10-7
Connecting and Hanging the Blood Transfer Bags
8. Remove the two blood transfer bags from the package
9. Connect the blood transfer bags to the sequestration lines as follows:
In PPP procedures, connect one of the blood transfer bags to the sequestration line with
the yellow clamp of the 4-way adapter. Close the clamp on the second sequestration line, as
it is not used.
In PRP1 procedures, connect one of the blood transfer bags to the sequestration line with
the orange clamp of the 4-way adapter. Close the clamp on the second sequestration line,
as it is not used.
In PRP2 procedures, connect the blood transfer bag for PPP collection to the sequestration
line with the yellow clamp and the other bag for PRP collection to the sequestration line
with the orange clamp. Close the clamp on the PRP sequestration line (orange). Make sure
the yellow clamp on the PPP sequestration line is open.
Figure 10-3 Blood Transfer Bag Setup: For PPP, PRP1, and PRP2 Procedures
10. Hang the blood transfer bags on the I.V. pole, keeping the inlet line face up (see Figure 10-4). To
prevent blood from accumulation in the upper parts of the bags, it is recommended to pass the
corresponding lines over the I.V. pole’s arm.
Figure 10-4 Hang the Blood Transfer Bags, Keeping the Inlet Line Facing Up