Operating Features of the Equipment
Centrifuge speed In ATS protocols: 5600rpm, accuracy ±40rpm
In SEQ protocols: PPP: 1500÷5600rpm, steps of 100rpm, accuracy ±50rpm
PRP1 and PRP2: 2400÷5600rpm, steps of 100rpm, accuracy ±50rpm
Peristaltic pump In ATS protocols the flow range is: 25÷1000mL/min, steps of 25mL/min
In SEQ protocols the flow range is: Fill/Spill phase: 10÷100mL/min, steps of
10mL/min
Empty phase: 25÷1000mL/min, steps of 25mL/min
Scale Capacity: 0 ˜ 5000 g
Accuracy: ± 5%
Hct indicator Type: optical
Position: Inlet/Outlet line
Range: 0% - 70%
Air sensor Type: ultrasonic waves ON / OFF with auto-test
Position: between pump and bowl
Waste line color indicator Type: optical
Position: waste line
Range: 0 - 3000 mg/dl
Waste bag level Type: Software
Intervention Range: 7 - 10 l
RBC sensor (or Buffy-Coat
sensor)
Low: optical linear CCD
High: punctual optical
Clamped empty line
sensor
Type: pressure sensor
Position: reinfusion line
Intervention: 1.5 - 3.0 bar
Centrifuge fluid loss
sensor
Type: resistive
Position: centrifuge well
Bar Code reader Type: optical
Position: top
Vacuum Pump (optional) In POST-OP mode the range is: -10 ÷ -100 mmHg, steps of -10 mmHg
In INTRA mode the range is: -50 ÷ -300 mmHg, steps of -10 mmHg
Accuracy: ± 20 mmHg intraoperative, ± 10 mmHg postoperative
Both POST-OP and INTRA aspiration modes are available with the Vacuum
module operating independently and operating connected to X
TRA.
Data storage Type: not-rechargeable buffer battery
Duration: (10 year functioning)
Communication ports 1 Ethernet (port not available in current software version)
3 RS232 serial
3 USB
1 RS422 serial
Table 16-2