EasyManua.ls Logo

sorin XTRA - Protocol Parameters (PPP and PRP)

sorin XTRA
304 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preoperative Sequestration (PPP and PRP)
10-24
20077/021 US XTRA Operators Manual
Empty Phase
The Empty phase of the PRP2 protocol, during which the remaining blood components in the bowl are
transferred to the RBC bag, is identical to that of the PPP and PRP1 protocols detailed in “PPP” on
page 10-22. Before the Empty phase begins, the operator will be prompted to close the PRP line and
open the PPP line so air can enter the bowl during emptying.
Protocol Parameters (PPP and PRP)
The table below contains the parameters recommended by LivaNova for the PPP and PRP protocols,
according to the bowl size used.
Note: Every time a new case is started, factory procedure parameters are reinitialized to their
default values, even if the unit is not shut down and restarted.
Protocol Parameters Range Step Unit Factory values
X/55 X/125 X/175 X/225
PPP Fill Flow 10 - 100 10 ml/min 20 50 50 60
Final Fill Flow
i
i
These parameters can be modified run time only and modifications will be lost at every new cycle.
10 - 100 10 ml/min 20 50 50 60
Fill Centrifuge Speed
ii
ii
These parameters, Centrifuge Speed in the Fill and in Spill phases, are related to each other: Spill Centrifuge Speed must always be lower
than Fill Centrifuge Speed.
1500 - 5600 100 rpm 5600 5600 5600 5600
Empty Flow 25 - 1000 25 ml/min 100 100 100 100
PRP1 Fill Flow 10 - 100 10 ml/min 20 50 50 60
Final Fill Flow
i
10 - 100 10 ml/min 20 50 50 60
Fill Centrifuge Speed
ii
2400 ˜5600 100 rpm 5600 5600 5600 5600
Spill PPP Flow 10 - 100 10 ml/min 10 10 10 40
Spill PPP Vol 0 - 400 1 ml 27 30 35 100
Spill PRP Flow
i
10 - 100 10 ml/min 10 10 10 20
Spill PRP Vol 0 - 400 1 ml 25 35 35 55
Spill Centrifuge Speed
ii
2400 - 5600 100 rpm 2400 2400 2400 2400
Empty Flow 25 - 1000 25 ml/min 100 100 100 100
PRP2 Fill Flow 10 - 100 10 ml/min 20 50 50 60
Final Fill Flow
i
10 - 100 10 ml/min 10 30 30 60
Fill Centrifuge Speed
ii
2400 - 5600 100 rpm 5600 5600 5600 5600
Spill PPP Flow 10 - 100 10 ml/min 10 10 10 40
Spill PPP Vol 0 - 400 1 ml 27 30 35 100
Spill PRP Flow
i
10 - 100 10 ml/min 10 10 10 20
Spill PRP Vol 0 - 400 1 ml 20 25 25 45
Spill Centrifuge Speed
ii
2400 - 5600 100 rpm 2400 2400 2400 2400
Empty Flow 25 - 1000 25 ml/min 100 100 100 100
Table 10-2 PPP and PRP Protocol Parameter Summary