Processing
X
TRA Operator’s Manual 20077/021 US 5-17
If an intraoperative protocol is interrupted and if the protocol set (defined in Configuration mode)
includes any Post-op protocols (factory or user defined), touching the "Retain" button will prompt the
following screen:
By touching “Retain Intra” the interrupted case will be resumed with the vacuum pump and protocol
switched to Intraoperative mode.
By touching “Retain Post-op” the interrupted case will be resumed with the vacuum pump and protocol
switched to Post-operative mode.
Dealing With Expected Warnings During a Cycle
This section describes the most likely warnings, along with the appropriate response to each, which
can be triggered during a processing cycle. For a complete description warnings and alarms, refer to
Chapter 14: Troubleshooting
.
Reservoir Empty
The Reservoir Empty warning is triggered during the Fill phase if air is detected in the reservoir line.
Also refer to “Reservoir empty. Bowl not filled” on page 14-20.
When triggered, the acoustic sequence is sounded, the message “Reservoir empty. Bowl not filled” is
displayed in the message area of the current screen, the reservoir displet is highlighted with a yellow
outline, and the processing cycle is paused (see Figure 5-23).
CAUTION
When retaining case data between power cycles, do not change the
disposable and/or its positioning to avoid potential errors during the
auto-load process.