Configuring XTRA
XTRA Operators Manual 20077/021 US 8-9
Volume of Alarm
Use this three-part toggle button at the top of the screen to set the volume for all acoustic alarms
(Low, Medium, or High).
Standard Tone
The tone of the standard audible alarms and beeps may be configured to one of three available
acoustic sequences.
Tone 1: three pulses, the last pulse is of a different tone
Tone 2: three pulses, the last pulse is of a different duration
•Tone 3: two pulses
Once selected, this sequence will be associated to all alarms/warnings that emit an endless acoustic
sequence.
Alarms and warnings that issue a finite acoustic sequence will always produce a fixed signal composed
of three short beeps.
Acoustic Signal
The duration of several alarms may be configured individually using each alarm’s two-part toggle
button under the “Acoustic signal” heading:
Long - continuously repeated acoustic alarm
Short - three single beeps
“RBC Bag full” Warning Counter Reset Option
By activating this option, it will be possible to reset the counter controlling the “RBC Bag full” warning.
Enable or Disable Bag Warnings
The buttons labeled “RBC bag full” and “Waste bag full” may be used to enable or disable the
corresponding bag warnings.
CAUTION
Any modification on the acoustic signals can make the operator take longer to
realize that the machine has made a warning.
CAUTION
The deactivation of the alarms “RBC bag full” and/or “Waste bag full” is under
the responsibility of the user who must directly control the fill level of the
bags.