Troubleshooting
X
TRA Operators Manual 20077/021 US 14-5
Memory not functioning.
Part of the data has
been lost.
Warning
Memory might have lost
user protocol names and/or
user protocol parameters.
The machine will continue
working anyway, but it is
recommended to call the
authorized technical
service.
Press X to clear the warning
message and continue
Memory not functioning
Empty clamp test failed
Warning
A hardware anomaly was
detected while testing the
Empty clamp.
Empty clamp test failed
Fill clamp test failed
Warning
A hardware anomaly was
detected while testing the
Fill clamp.
Fill clamp test failed
HCT indicator
calibration failed
Warning
The machine cannot
calibrate the HCT indicator.
Verify that the line is
properly seated in the
indicator and not dirty on
the outside. Press PLAY to
retry (max. 3 attempts)
After 3 failures, press
DELAY CALIBRATION to
exit. The system can still
be operated, but the HCT
indicator will not be
available for the current
case.
If the problem persists, call
the authorized technical
service.
HCT indicator calib. failed
Message Text
Warning
or Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response Data Management Text
Table 14-1 List of Warnings Which Might Occur During Setup (Continued)