Troubleshooting
X
TRA Operator’s Manual 20077/021 US 14-7
Pump loop loading
failed
Warning
The machine cannot
properly load the pump
loop.
• If the pump loop is
blocking the rotor,
manually remove it.
Check for any eventual
damage to the pump loop
and, if present, replace
the disposable. If the
pump loop is not
damaged, manually seat
it and press PLAY to
resume.
• If the rotor is blocked by
any other object, remove
it, then press PLAY to
resume.
If the problem persists, call
the authorized technical
service
Pump loop loading failed
Pump rotor out of
position
Warning
The pump rotor cannot
rotate, it might be missing
or not seated correctly.
1. Open the centrifuge lid
and check the position
of the pump rotor:
• open the white lever
• push the pump rotor fully
down
• turn the pump rotor to
align the white lever with
the slot in the metal shaft
• close the white lever into
the slot
• make sure the white level
is completely down
2. Close the centrifuge lid
and press PLAY to
resume.
If the problem persists, call
the authorized technical
service
Pump rotor mispositioned
Wash clamp test failed
Warning
A hardware anomaly was
detected while testing the
Wash clamp.
Wash clamp test failed
Message Text
Warning
or Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response Data Management Text
Table 14-1 List of Warnings Which Might Occur During Setup (Continued)