Troubleshooting
14-12
20077/021 US XTRA Operator’s Manual
Bowl should be full
Alarm
Air has been detected while
extracting blood from the bowl
but the bowl should be full.
• Press FAT REM to resume the
Fat Removal phase.
• Press EMPTY to switch to the
Empty phase.
WARNING: Skipping the Fat
Removal phase now may result
in a dilution of the processed
blood.
If the problem persists, call the
authorized technical service.
Bowl should be full
Bowl not filled / not
washed
Alarm
The current phase has not been
finished yet.
From a processing screen:
• Press FILL to resume the Fill
phase.
• Press WASH to start / resume
the Wash phase.
WARNING: Inadequate filling
may lead to lower than
expected hematocrit and
removal of waste components.
• c) Press EMPTY to confirm the
Empty phase.
WARNING: Inadequate washing
of packed red blood cells may
lead to an excessive level of
contaminants (e.g.,
anticoagulant, plasma free
hemoglobin) in the processed
blood.
From the End of Case Screen:
• Press X to exit and return to
End Case screen.
•Press EMPTY to confirm the
Empty phase.
WARNING: Inadequate filling
may lead to lower than
expected hematocrit and
removal of waste components.
Inadequate washing of packed
red blood cells may lead to an
excessive level of contaminants
(e.g., anticoagulant, plasma
free hemoglobin) in the
processed blood.
Bowl not filled/washed
Message Text
Warning or
Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response
Data Management
Text
Table 14-2 List of Warnings and Alarms Which Might Occur During Processing (Continued)