Troubleshooting
14-24
20077/021 US XTRA Operators Manual
Electronic Failure - Memory or
Microcontroller
E3
E16
E26
E27
E28
E30
E36
E40
E61
E87
Switch the machine OFF.
Wait a few seconds, switch the machine on and
touch the “Retain” button.
Electronic Failure - Serial Communication
E4
E5
E6
E8
E9
E10
E12
E13
E14
E31
E32
E39
Switch the machine OFF.
Wait a few seconds, switch the machine on and
touch the “Retain” button.
Electronic Failure - NMC Board
Overtemperature
E23
Switch the machine OFF.
Wait a few minutes to allow the board to cool
down, then switch the machine on and touch the
“Retain” button.