XTRA Operator’s Manual 20077/021 US A-3
Test
frequency
(MHz)
Band
(MHz)
Service Modulation
Maximum
power
(W)
Distance
(m)
Immunity test
level
(V/m)
385 380 – 390 TETRA 400
Pulse modulation
18 Hz
1.8 0.3 27
450 430– 470
GMRS 460
FRS 460
FM
± 5 kHz deviation
1 kHz sine
20.328
710
704 – 787 LTE Band 13, 17
Pulse modulation
217 Hz
0.2 0.3 9745
780
810
800 – 960
GSM 800/900
TETRA 800
iDEN 820
CDMA 850
LTE Band 5
Pulse modulation
18 Hz
20.328
870
930
1720
1700 – 1990
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE Band 1, 3, 4,
25
UMTS
Pulse modulation
217 Hz
20.328
1845
1970
2450 2400 – 2570
Bluetooth
WLAN 802.11 b/
g/n
RFID 2450
LTE Band 7
Pulse modulation
217 Hz
20.328
5240
5100 – 5800 WLAN 802.11 a/n
Pulse modulation
217 Hz
0.2 0.3 95500
5785
Note: If necessary to achieve the immunity test level, the distance between the transmitting antenna and the X
TRA System
may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
Note: Portable radio frequency (RF) communications equipment (including peripherals such as antenna cables and external
antennas) as well as cables specified by the manufacturer should be used no closer than 30 cm to any part of the X
TRA
System. Otherwise this could result in performance degradation of this equipment.
Table A-3 Test specifications for enclosure port immunity to RF wireless communications equipment