Bad Read on This error can occur in Length/Strength
Track Home Calibration, Lens-to-Break, and Length/Strength
Module Testing. The MC4 Motor Controller failed
to detect that the comb was in the home position.
Bad Sample
When the calculated values are out of tolerance, a
tone is sounded, and a new sample must be tested.
Reset the instrument, then recalibrate.
Bad Send (MC4)
The computer timed-out while waiting for
the motor controller to respond to a
signal.
Breaker Non-
The controlling program has issued a
Operational breaker request, but no data resulted from the
request.
Check Force
When the force is not within +1.0 and -2.0, reset
the motors. If this does not resolve the problem,
call a Service Technician.
Check Lamp
The lamp voltage is outside the range
Voltage established in the Setup utility.
Color Limit
When the measured value is outside the lot limit (if
Lot Limits are YES on the 900 Status Menu), an
instruction statement is displayed to allow you to
accept or reject the value. If you reject the value,
repeat the test on the sample.
Check Optics
When the optics are not within ±10, turn the
recessed knob on the inside lip of the left console
(near the middle) until the message disappears and
the flashing numbers are close to 0.
Check Vacuum
The vacuum voltage is outside the range
Manifold established in the Setup utility.