Chapter 14
Version 4.0 (1.99) Appendix - Page 14-37
Color Done The color button was pressed after the color
reading was taken for the current sample.
Door Open
A micronaire reading was initiated, yet the
micronaire door was open (or the door sensor was
not engaged).
Error Reading
The MC4 Motor Controller failed to detect that
Brush Home the brusher was in the home position.
Force
The force reading is outside the
valid range (-2 to +1).
Large Sample W
hen the sample beard is too large to meet the
limits set during the initial instrument setup, prepare
another sample, and repeat the test.
Len Count
Too many length interrupts were-
generated by the motor controller.
Len Error
Length calculation resulted in a value
less than zero or greater than 99.9.
Mic Error
A micronaire value entered from the
keyboard was outside the valid range
(2.0 to 7.0).
Mic Limit
When the measured micronaire value is
outside the lot limit, this message will
display. It is also displayed if there is an
air leak in the chamber. Contact an
Uster Technologies Service Technician.
Motor Controller
This can occur during any phase of the
Not Responding operation if the MC4 Motor Controller is
not functioning properly.