Motors A problem has developed with the
synchronization between the 900
HVI computer and the motor
controller.
No Motor Control
A problem has developed with the
synchron- ization between the 900 HVI
computer and the motor controller.
No Motors
A problem has developed with the
synchronization between the 900 HVI
computer and the motor controller.
Optics
The optics reading is outside the valid
range (-10 to +10).
Out of Tolerance
This message occurs if the calculated
slope and offset are outside the given
range during micronaire calibration.
Perform calibration again. If the
message reappears, zero the air system,
and set the slope and offset to standard
values. Perform calibration again. If the
message reappears again, contact a
Uster Technologies Service Technician.
Printer Not Ready
The instrument is attempting to write to
the printer while the printer is OFF, out of
paper, off-line, or malfunctioning. This
message is displayed only if the status
parameters indicate that data are to be
sent to a printer. Check to make sure the
printer is "On-Line" and retransmit.
R.H.
The relative humidity, as reported by the
optional probe, is outside the range
specified in the Status Menu.