3.4 Functions Selected from the Main Menu
HA-8190V 3-9
[Meas. condition menu] screen (3000)
[Meas. result menu] screen (4000)
Option Description (default bolded) Page
User coefficient
setup (3100)
Sets coefficients “a” and “b” of the correction formula, “Y=aX+b”, for obtaining
HbA1c and HbF values in the Fast and Variant modes, respectively.
3-13
HbA1c (mmol/mol) a:0.01 to 1.50 (1.00), b:-50.0 to 50.0 (0.0)
HbF (%) a:0.01 to 3.00 (1.00), b:-5.00 to 5.00 (0.00)
Measurement No.
setup (3200)
Configures the measurement numbering system. 3-14
Meas. start No. reset: Power ON, No reset, Every batch
Meas. No. assignment: Sequential order, Port No.
Meas. condition
setup (3300)
Sets HbA1c control measurement parameters, HbA1c calibration conditions,
pressure unit for the column and error trigger conditions.
3-16
Ctrl expected value mmol/mol L: 0 to 99 mmol/mol (0 mmol/mol)
H: 0 to 200 mmol/mol (0 mmol/mol)
Ctrl measurement count: 1 to 3 times (3 times)
Control error range mmol/mol L: 0 to 99 mmol/mol (3 mmol/mol)
H: 0 to 99 mmol/mol (4 mmol/mol)
Action on control error: Stop measurement, No action, Issue warning
Pressure unit: kg/cm
2
, MPa
STD. solution meas.: 1 to 3 times (3 times)
Calibration error range (%): 0.0 to 9.9% (3.0%)
Replace column notification: ON, OFF
Calibration notification: ON, OFF
Misread barcodes: 0 to 150 times (0 time)
No. of failed smpl tube spin: 0 to 150 times (0 time)
Eluent CV pack: ELUENT 90CV (600mL),
ELUENT 90CV-S (200mL)
Option Description (default bolded) Page
Print (4100) Prints out measurement results and chromatograms. 3-18
Date (DD-MM-YY): Range of measurement dates (01-01-2000 to 31-12-2099)
Measurement: Normal & STAT meas., Normal measurement,
STAT measurement, Control measurement, All
Results: All, Normal results only, Include abnormal results,
Barcode misread
Range of results: All, Meas No., Port No., Sample ID
Meas. mode: Variant, Fast, Fast & Variant (Current measurement mode)
Print list (4200) Prints out a list of measurement results.
* For information on the setup items, see [Print] above.
Transmit (4300) Transmits measurement results to the external device.
* For information on the setup items, see [Print] above.
View (4400) Displays measurement results stored in the memory of the instrument. You can
also edit sample IDs of measurement results and make settings for control
measurement results.
* For information on the setup items, see [Print] above.
3-21
Delete (4500) Deletes measurement results and history of warnings, errors and troubles from the
memory.
3-24
Items to delete: All, Normal & STAT meas., Control measurement,
Warning/Error/Trouble