3.5 Measurement Conditions
HA-8190V 3-17
Setup item Description (default bolded)
Ctrl expected value
mmol/mol
Set control expected values of the controls you are using. These values are written on the
package insert of the controls.
L: Low solution 0 to 99 mmol/mol (0 mmol/mol)
H: High solution 0 to 200 mmol/mol (0 mmol/mol)
Ctrl measurement
count
Set the number of times for measuring the control in a single sample container. Up to 3
measurements can be performed per sample container, so you can reduce the volume of
controls used for control measurements.
Range: 1 to 3 times (3 times)
Control error range
mmol/mol
Set the error detection range used in control measurements. The instrument issues a warning
(W-011 or W-071) if the difference between the obtained control measurement result and
<Control expected values> exceeds the values set here.
L: Low solution 0 to 99 mmol/mol (3 mmol/mol)
H: High solution 0 to 99 mmol/mol (4 mmol/mol)
Action on control
error
Select the action to take place when the difference between the obtained control
measurement result and <Control expected values> exceeds <Control error range>.
Stop measurement: Issues W-071 and stops measurements.
No action: Continues measurements without issuing a warning.
Issue warning: Issues W-011, but continues measurements.
Pressure unit Select the pressure unit for the column.
Options: kg/cm
2
, MPa
STD. solution meas. Set the number of times to measure each of the Low and High solutions for automatic
calibration.
Range: 1 to 3 times (3 times)
Calibration error
range (%)
Set the error detection range used in automatic calibration.
Range: 0.0 to 9.9% (3.0%)
An error will occur if the measured values obtained respectively with the Low and High
solutions produce the following condition.
Example: When “3.0%” is set
| Measured value – Mean value | > Mean value 3.0%
* The above “mean value” refers to the mean value of the measured values
obtained with the Low or High solutions.
Replace column
notification
Set whether or not to notify the user by a message that the column requires replacement
when the column counter exceeds a preset number.
Options: ON, OFF
Calibration
notification
Sets whether or not to notify the user that calibration is required when the power is turned on.
This notification is issued when calibration was not performed for both the Fast and Variant
modes at the same time or was performed in one of the modes only.
Options: ON, OFF
Misread barcodes Set the number of misread barcode errors in a batch at which the instrument issues W-081
and stops measurement. If set to “0”, measurement does not stop regardless of times the
error occurs.
Range: 0 to 150 times (0 time)
No. of failed smpl
tube spin
Set the number of sample tube spin failures in a batch at which the instrument issues W-082
and stops measurement. If set to “0”, measurement does not stop regardless of times the
failure occurs.
Range: 0 to 150 times (0 time)
Eluent CV pack Selects the type of eluent CV pack to be used.
Options: ELUENT 90CV (600mL), ELUENT 90CV-S (200mL)