EasyManua.ls Logo

Arkray ADAMS A1c - Page 155

Arkray ADAMS A1c
288 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Auxiliary Operations
HA-8190V 3-40
For sample cups:
Odd-numbered ports: Empty
Even-numbered ports: Sample cup
d Load the hemolysis control rack onto the sampler.
2 Measure the samples.
a Touch .
Measurement will start.
A result will be displayed for each measurement.
If “Measurement ended.” appears:
The most recent result will appear on the display.
If an error or trouble message appears:
Trouble was detected. See the relevant page in “Chapter 5
Troubleshooting” to take the appropriate action.
b Touch and then to return to the standby
screen.
c Check that the hemolysis control rack is not moving, and remove the rack from the rack unloading side of the
sampler.
Sample cup with
hemolysis sample
Empty
12345678910
REFERENCE:
To stop measurement, touch or .
Start
Stop
Go back
HOME

Table of Contents

Related product manuals