4.2 Replacement of Consumables
HA-8190V 4-25
5 Remove air from the tube by priming.
a Perform priming for the tube and pump fluid to it.
Eluent nozzle: Perform priming for the applicable eluent, and then pump fluid to the tube.
Hemolysis washing solution nozzle: Perform priming for hemolysis washing solution.
See “4.4. Priming” on page 4-52.
b Check that air has not entered the tube from the filter holder.
If there is air in the tube, further tighten the push screw that was tightened in step 2-e with a wrench.
If air still enters the tube, retry from step 1-a.
REFERENCE:
You can record the date of mesh filter replacement on the instrument. See “4.5.1. Recording the Date of
Maintenance” on page 4-53.