EasyManuals Logo

Arkray ADAMS A1c User Manual

Arkray ADAMS A1c
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #249 background imageLoading...
Page #249 background image
Chapter 5 Troubleshooting
HA-8190V 5-20
T-610 - T-612 Measurement-side lever drive
Cause Sample racks are improperly loaded onto the rack loading side of the sampler.
Something is obstructing the path of the measurement-side lever on the sampler.
The measurement-side lever did not operate properly.
Remedy a Touch to clear the trouble.
b Load the sample racks onto the rack loading side properly.
c Remove the obstruction from the path of the measurement-side lever located on the instrument
side of the sampler.
d If the trouble persists, turn off the power and contact your distributor.
T-613, T-614 Return-side lever drive
Cause Sample racks are improperly loaded onto the rack unloading side of the sampler.
Something is obstructing the path of the return-side lever on the sampler.
The return-side lever did not operate properly.
Remedy a Touch to clear the trouble.
b Load the sample racks onto the unloading side properly.
c Remove the obstruction from the path of the return-side lever located on the front side of the
sampler.
d If the trouble persists, turn off the power and contact your distributor.
T-620 - T-622 Sample tube spin F/B
Cause Something is obstructing the sample tube spinning unit.
The sample tube spinning unit did not operate properly.
Remedy a Touch to clear the trouble.
b Remove the obstruction from the sample tube spinning unit (towards the rear under the STAT
port cover).
c If the trouble persists, turn off the power and contact your distributor.
T-623 Sample tube spinning
Cause The sample tube spinning unit did not operate properly.
Remedy a Touch to clear the trouble.
b Turn off the power and contact your distributor.
T-800
T-805
Serial transmission
Serial receiving
Cause The serial communication board did not operate properly.
Remedy a Touch to clear the trouble.
b Turn off the power and contact your distributor.
T-801 - T-804
T-807 - T-810
Two-way communication
Cause The communication cable of the external device is disconnected or improperly connected.
Remedy a Touch to clear the trouble.
b Properly connect the communication cable.
c If the trouble persists, turn off the power and contact your distributor.
T-811 No matching ID
Cause The host computer transmitted an abnormal measurement command.
Remedy a Touch to clear the trouble.
b Check if the host computer correctly replies to an inquiry from the instrument.
OK
OK
OK
OK
OK
OK
OK

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray ADAMS A1c and is the answer not in the manual?

Arkray ADAMS A1c Specifications

General IconGeneral
BrandArkray
ModelADAMS A1c
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals