Chapter 5 Troubleshooting
HA-8190V 5-26
b Pull the fuse holder straight to the front.
3 Replace the fuse.
a Remove the blown fuse from the fuse holder and set a new fuse.
4 Store the fuse holder.
a Insert the holder into its original position.
b Close the fuse holder cover.
5 Turn on the power.
a Plug the power cord into the power input terminal on the rear panel.
b Plug the power cord into an outlet.
c Press the main power switch on the rear panel to turn on the main power.
d Press the standby switch to turn on the power.
The standby switch will light up.
Fuse holder
New fuse
NOTE:
Align the arrows on the fuse holder and cover, and insert the
holder into place.